Detectors for high spatial resolution dosimetry and micro-dosimetry in proton and heavy ion therapy

17 Mar 2021, 11:30
30m
Budker INP

Budker INP

Lavrentiev av. 11, Novosibirsk, Russia
WG3: Accelerator technologies for industrial & medical applications WG3

Speaker

Prof. Anatoly Anatoly Rosenfeld (Centre for Medical Radiation Physics , School of Physics, University of Wollongong , Australia)

Description

Based on many years of experience in development of silicon-on-insulator (SOI) microdosimeter, the Centre for Medical Radiation Physics, University of Wollongong, has successfully developed a microdosimetric probe which is based on a SOI microdosimeter with 3D micron sized sensitive volumes (SVs) array mimicking dimensions of cells, known as the “MicroPlus-Mushroom” microdosimeters, to address the shortcomings of the tissue equivalent proportional counter (TEPC)
A method for converting silicon microdosimetric spectra to tissue for a therapeutic proton and heavier ion beams, based on Monte Carlo simulations was developed. The MicroPlus microdosimeter provides extremely high spatial resolution and were used to evaluate the relative biological effectiveness (RBE) of 12C, 14N 16O , 56Fe, 20Ne ions at Heavy Ion Medical Accelerator in Chiba (HIMAC), Japan as well as to measure the microdosimetric distributions of a proton pencil-beam scanning (PBS) and passive scattering system at different proton therapy centres. Good agreement between predicted cell survival response using MKM and measured from in vitro experiments in the same radiation field allow replacing time consuming cell experiments with MicroPlus microdosimeter measurements.
Another application of SOI microdosimeter is for evaluation of radiation shielding and radiation protection of astronauts in radiation environment typical for SPE and GCR. We demonstrated that SOI microdosimeters are suitable for in situ evaluation of radiation shielding efficiency of multi-layered space craft and astronaut shelter walls in radiation fields on accelerators mimicking SPE and GCR. SOI microdosimeters supplement Monte Carlo simulation which not always accurate due to lack of knowledge of cross sections and time consuming.

Silicon detectors for fast verification of range, energy and spot position of proton and carbon ion pencil beam scanning for daily QA will be presented.

Primary authors

Prof. Anatoly Anatoly Rosenfeld (Centre for Medical Radiation Physics , School of Physics, University of Wollongong , Australia) Dr Linh Tran (Centre for Medical Radiation Physics , University of Wollongong) Prof. Susanna Guatelli (CMRP) Dr David Bolst (CMRP) Dr Emily Debrot (CMRP) Mrs Stefania Peracchi (CMRP) Mr Ben James (CMRP) Mr Vladimir Pan (CMRP) Mr James Vohradsky (CMRP) Prof. Marco Petasecca (CMRP) Prof. Michael Lerch (CMRP) Dr Dale Prokopovich (EBG MedAustron) Dr Marco Povoli (SINTEF) Dr Angela Kok (SINTEF) Dr Taku Inaniwa (NIRS) Prof. Naruhiro Matsufuji (NIRS)