SUMMARY:Study of  cryogenic photomultiplier tubes for the future double-ph
 ase cryogenic avalanche detector.
DTSTART;VALUE=DATE-TIME:20170228T100000Z
DTEND;VALUE=DATE-TIME:20170228T110000Z
DTSTAMP;VALUE=DATE-TIME:20261108T085252Z
UID:indico-contribution-1628@indico.inp.nsk.su
DESCRIPTION:Speakers: Lev Shekhtman (Budker Institute of Nuclear Physics)\
 nWe report the results of characterization study of several types of cryog
 enic photo-multipliers manufactured by Hamamatsu Photonics\, namely: compa
 ct 2 inch R6041-506MOD tubes\, 3 inch R11065-10 and R11065-MOD tubes for o
 peration in liquid Ar and 3 inch R11410-20 tubes for operation in liquid X
 e. These types of PMT are proposed for installation into the future double
 -phase cryogenic avalanche detector that is developed in the Laboratory of
  Cosmology and Particle Physics of the Novosibirsk State University jointl
 y with the Budker Institute of Nuclear Physics. \nCompact PMTs are planned
  to be installed at the side of the cryogenic vessel and they will detect 
 electroluminescense in a high field region above liquid. 3 inch PMTs will 
 be installed at the bottom of the cryogenic vessel in liquid Ar and they w
 ill detect primary scintillations from recoil nuclei as well as secondary 
 scintillations. Main task  of both PMT systems is to provide fast trigger 
 signal. Therefore the PMTs have to distinguish between single electron and
  double electron signals to suppress effectively the background. \nEight R
 11065 PMTs and eight R11410-20 tubes were tested and they all demonstrated
  excellent performance in terms of gain and relative single electron effic
 iency. All 3 inch PMTs showed maximal gain in liquid Ar above 5x10$^6$ and
  relative single electron efficiency above 95%.  Compact R6041-506MOD tube
 s have different dynode  system and thus their single electron energy reso
 lution and relative efficiency is much worse than that of 3 inch tubes. Fr
 om 21 two inch PMTs only 12 tubes were selected with acceptable relative s
 ingle electron efficiency higher than 70% at the maximal gain higher than 
 5x10$^6$. However these PMTs are very attractive because these are the onl
 y compact type of tubes that can operate in liquid Ar.\n\nhttps://indico.i
 np.nsk.su/event/8/contributions/1628/